Call us at 732 292-1994 or 800 867-6690 | Email us at triadscientific@gmail.com | Visite us on
PLANET BEST
Lab Instrumentation
Lab Equipment
Lab Furniture
Yamato Spray Dryer ADL311SA S Type is for Organic Solvent NEW
Overview:
"Yamato’s ADL-311S does not require the liquid sample or solution to undergo any pre- or post-processes such as filtration, separation, or pulverization. The use of organic solvents is fully supported with the attachment of our GAS-410 organic solvent recovery unit. Small, expensive and/or heat sensitive samples can be dried quickly and efficiently with this easy to operate system. " (Yamato)
Advantages:
Processes samples as small as .5 g of solid matter
Safe for heat-sensitive samples, such as food or medical products
No risk of contamination
Digital display of inlet/outlet temperature and drying air volume increases readability
Detachable drying chamber, cyclone and product vessel
Fast and easy clean up
Universal power supply and multilingual touch screen controller
Specifications:
|
Performance |
Water Evaporation Rate | Max. approx. 1,300 ml/h |
Temperature Control Range | 40˚C - 220˚C at inlet | |
Temperature Control Accuracy | ± 1°C at inlet | |
Dry Air Flow Rate | Max. 0.7 m³/min | |
Air Spray Pressure Control Range | 0 - 249 k Pa (0-3 kg/cm²) | |
Flow Rate for Sample Liquid Feed Pump | Max. 26 ml/min | |
Structure |
Temperature Controller | PID digital tempertaure controller |
Temperature Sensor | K thermocouple | |
Stainless Tube Heater | 2.88 kW at 240V | |
Sample Liquid Feeding Pump | Quantitative peristaltic pump | |
Solvent Recovering Capability | Organic solvent recovery unit GAS-410 must be used | |
Spray Line Cleaning | Nozzle orifice cleaning (manual pulse-jet system) | |
Safety Devices | Self-diagnostic functions (e.g. temperature aberration); Sample feed reversal | |
Standard |
Air Spray Pressure Gauge | Bourdon tube: 294k Pa (3 kg/cm²) |
External Dimensions (WxDxH) | 23.7 x 23.3 x 53.2 in | |
Weight | 125 kg or 275.6 lbs (including GF-300) | |
Power Supply | AC 200V - 240V, single-phase 18 A | |
Included Accessories |
Sample liquid tube | Silicone tube - 2 pcs |
Safety Cover | Yes | |
Air hose | 1 pc | |
Exhaust Duct | 1 pc | |
Included Accessories |
Mini Spray Attachment | GF - 300 |
Optional Accessories |
Organic Solvent Recovery Unit | GAS - 410 |
Inlet/outlet Temperature Recorder | 212792 - factory installed | |
Fine Powder Recovery Cyclone | 212780 | |
Viton/Tiron Feeding Tube | W0180001 | |
Nozzle | 1A, 1, 2A, 2, 3 | |
Operational Requirements |
Compressed Air | 28 l/min or 3 kgf/cm²; not required when using GAS-410 |
Type of Gas | N₂ gas (99.99% purity, medical grade) is required when using GAS-410 |
Applications: Pharmaceuticals, Food and beverage, Cosmetics and fragrances, Advanced ceramics, Chemical and material science, Agricultural Energy
For extensive product details, visit the product page at http://www.yamato-usa.com/products/spray/adl311sa.html