Call us at 732 292-1994 or 800 867-6690 | Email us at triadscientific@gmail.com | Visite us on
PLANET BEST
Lab Instrumentation
Lab Equipment
Lab Furniture
Yamato Laboratory Spray Dryer series DL410(CE) NEW
Overview:
"The DL-410 is a larger capacity spray dryer that also does not require the liquid sample or solution to undergo any pre or post-processes such as filtration, separation, or pulverization. The use of organic solvents is fully supported with the attachment of our GAS-410 organic solvent recovery unit. Small, expensive and/or heat sensitive samples can be dried quickly and efficiently with this easy to operate system. " (Yamato)
Advantages:
Processes samples as small as .5 g of solid matter
Safe for heat-sensitive samples, such as food or medical products
No risk of contamination
Digital display of inlet/outlet temperature and drying air volume increases readability
Detachable drying chamber, cyclone and product vessel
Fast and easy clean up
Universal power supply and multilingual touch screen controller
Easy operation and maintenance
The hot air inlet and drying chamber cover automatically move up and down, and since the cyclone and product
vessel can easily be removed, cleaning and maintenance after your experiment is easy.
Control functions are conveniently arranged on the control panel for various conditions.
The temperature recorder, air flow meter, pressure gauge and other measurements allow easy control of experiment conditions.
Specifications:
Performance | Water Evaporation Rate | Max. approx. 3,000 ml/h |
Temperature Control Range | 40˚C - 300˚C at inlet | |
Temperature Control Accuracy | ± 1°C at inlet | |
Dry Air Flow Rate | Max. 1.2 m³/min | |
Air Spray Pressure Control Range | 0 - 600 k Pa (0-6 kg/cm²) | |
Flow Rate for Sample Liquid Feed Pump | Max. 80 ml/min | |
Structure | Temperature Controller | PID digital temperature controller |
Temperature Sensor | K thermocouple | |
Stainless Tube Heater | 2kW+2kW at 240V | |
Sample Liquid Feeding Pump | Quantitative peristaltic pump | |
Solvent Recovering Capability | Organic solvent recovery unit GAS-410 must be used | |
Spray Line Cleaning | Needle inside the nozzle to clean the mesh automatically | |
Safety Devices | Self-diagnostic functions (e.g. temperature aberration); Sample feed reversal | |
Standard | Air Spray Pressure Gauge | Bourdon tube: 600k Pa (6 kg/cm²) |
External Dimensions (W x D x H) | 1750 x 1060 x 880 mm or 69 x 42 x 35 in | |
Weight | 180 kg or 397 lbs | |
Power Supply | AC 200V - 240V, single-phase 24 A | |
Included Accessories | Sample liquid tube | Silicone tube - 2 pcs |
Safety Cover | Yes | |
Air hose | 1 pc | |
Exhaust Duct | 1 pc | |
Required Accessories | Mini Spray Attachment | N/A |
Optional Accessories | Organic Solvent Recovery Unit | GAS - 410 |
Inlet/outlet Temperature Recorder | 212792 - factory installed | |
Fine Powder Recovery Cyclone | N/A | |
Viton/Tiron Feeding Tube | Please inquire | |
Nozzle | 3, 4, 5 | |
Operational Requirements | Compressed Air | 28 l/min air volume and 8kgf/cm2 compressed air is required |
Type of Gas | N₂ gas (99.99% purity, medical grade) is required when using GAS-410 |