Hiranuma Coulometric Karl Fischer Titrator AQ-2200A
BRAND NEW with ONE YEAR FACTORY WARRANTY
FEATURES:
1. Intuitive control panel display!
Current status, data and function keys are displayed on a colorful and large screen. Color touch screen is 7.5-inches wide. All helpful and necessary information is always at hand!
2. Browse data by PC on a LAN connection!
Result data with curves can be viewed on your PC internet browsing program without optional software.
3. Enhanced memory data capacity!
Store results data on the USB memory stick and the stored data can be read out on the color display.
4. Built-in thermal printer!
Easy paper loading. High resolution, clear printout.
5. Two KF channels run in parallel!
"Combining one additional KF station with either a volumetric or coulometric module enables two different measurements to be performed at the same time. With both a volumetric and coulometric channel, one can measure moisture over the entire range from a few ppm to 100% or double your sample throughput with 2 identical KF stations." (HIRANUMA)
SPECIFICATIONS:
ITEM | SPECIFICATIONS |
---|---|
Titration method | Coulometric Karl Fischer Titration |
Multitasking function | Up to two KF titration stations (Double coulometric channels or combination of coulometric and volumetric channels) |
End-point detection method | AC polarization potentiometric detection method |
Background correction | Automatic correction (selectable ON/OFF) |
Measuring range | 10μgH20 - 99mgH2O |
Display unit | μgH2O,ppm,% |
Accuracy | Within CV of 0.3% (1g of 1mgH2O/mL water-in-methanol measured 10 times) |
Required measurement time | FAST :1.8mgH2O/min(30μgH2O/sec) MEDIUM:1.2mgH2O/min(20μgH2O/sec) SLOW :0.6mgH2O/min(10μgH2O/sec) |
Electrolytic cell capacity | 150mL (min.usage: approx. 100mL) |
Memory data capacity | Internal memory: 100 data / channel USB flash memory (≥256MB): Max. approx. 9900 data |
GLP compliant functions | Electrolytic cell maintenance, Measurement accuracy test |
Calculation modes | 6 modes corresponding to sample status and sampling methods |
Printer | Built-in thermal printer (paper width: 58mm) *Optional external printer: Dot impact printer PR-302B |
Touch panel screen | 7.5 inch color graphics LCD |
External input/output |
RS-232C: 1 port, LAN: 1 port, USB flash memory stick: 1 port |
Power supply | AC100-240V, 50/60Hz, 100VA |
Dimensions / Weight | Main unit: 220(W) x 400(D) x 200(H) mm / Approx. 5kg |