Jenway 3540 Combined pH and Conductivity Meters Jenway 3540 Enterprise Conductivity/pH Meter Jenway Conductivity meter 3540 Jenway pH meter 3540
• Simultaneous display of pH, conductivity and temperature
• Automatic or manual calibration
• Combined pH, conductivity and temperature probe
• Power on self check routine
• Auto-selecting ranges for both TDS and conductivity modes
3540 Bench Combined Conductivity/pH Meter
• Simultaneous display of pH and conductivity
• 2 independent channels to prevent interference between probes
• Additional modes for resistivity, salinity, TDS and mV
• Data logger with auto save/print options
• Storage of up to 500 readings (250 for each mode)
The 3540 is ideal for use in all laboratories where both pH and conductivity analysis is required. The setup menu gives quick and easy access to the whole range of instrument, pH and conductivity measurement options. Calibration of both channels is automatic, with the option to select 1, 2 or 3 calibration points. The pH channel can also display mV values if required.
Technical Specification
Parameter 3540
pH range -2.000 to 20.000
pH resolution 0.001/0.01/0.1
pH accuracy ±0.003
Auto. buffer recognition Jenway, NIST, DIN and JIS
mV range - ±1999.9mV
mV resolution - 0.1mV
mV accuracy - ±0.2mV
Conductivity ranges Auto selected from 0.01uS to 1.999S
Conductivity resolution 0.01uS to 1mS
Conductivity accuracy ±0.5% ±2 digits
Cell constant 0.010 to 19.999
Automatic standard recognition 10uS, 84uS, 1413uS, 12.88mS (with manual override)
TDS range 6 auto selected from 0 to 1999g/l
TDS resolution 0.01mg/l to 1g/l
TDS accuracy ±0.5% ±2 digits
EC ratio (TDS) 0.50 to 0.80
Temperature range -10 to 105°C
Temperature resolution 0.1°C
Temperature accuracy ±0.5°C
ATC and manual 0 to 100°C
Temperature coefficient 0.00 to 4.00% / °C
Reference temperature 18, 20, 25°C
Additional modes - Salinity and resistivity
GLP - Calibration reminder interval, operator and sample ID and security coded setup
Outputs - Analogue, RS232 and IrDA interface
Connector BNC (pH) and 7 pin DIN (conductivity)
Power 9VAC (adaptor included)
Size (l x w x d) 210 x 250 x 55mm / 7 x 3 x 1.4 in
Weight 850g / 2lbs