GL Sciences LINer EXchanger (LINEX) NEW
"Automatically changing liners is a technique that was introduced by ATAS GL into the analytical market in the year 2001. The LINEX is our second generation liner exchanger based on the CTC PAL systems. The LINEX is an option of the OPTIC Multi Mode Inlet system.
LINEX can be used for Direct (in-injector) Thermal Desorbtion (DTD). In the automated mode liners are transported between tray and inlet by the CTC PAL injection systems equipped with a gripper. Any of the standard OPTIC liners can be handled by LINEX.
With LINEX, the multi-sample analysis sequence works in a simple way: head of the injector is opened and a liner containing sample is introduced. The head is closed and the liner is purged with carrier gas. Next, the injector is heated and the sample is transferred onto the column. At the end of the analysis the empty liner is moved back to the tray and the cycle is repeated.
LINEX can be used for the sample introduction of both solid and liquid samples. Also sample tubes filled with Tenax can be used for thermal desorption techniques. There is also a possibility to apply the approach for the analysis of samples with a difficult matrix using the so-called DMI technique. With DMI a minimum sample preparation is required." (GL)
FEATURES:
Brochure: http://www.glsciences.eu/liner-exchanger/LINEX.pdf